3 models are now on 15% sale.
Also get a free HART Drywell cable with every Fluke 9142, Fluke 9143 or Fluke 9142 purchase!
9142/9142-P
|
9143/9143-P
|
9144/9144-P
|
|
---|---|---|---|
Range
(23 °C amb.)
|
–25 °C to 150 °C
|
33 °C to 350 °C
|
50 °C to 660 °C
|
Base unit display accuracy
|
± 0.2 °C (full range)
|
± 0.2 °C (full range)
|
± 0.35 °C at 420 °C
± 0.5 °C at 660 °C |
Process version accuracy
|
± 0.01 °C at –25 °C
± 0.025 °C at 150 °C |
± 0.02 °C at 50 °C
± 0.04 °C at 350 °C |
± 0.02 °C at 50 °C ± 0.07 °C at 660 °C |
Stability
|
± 0.01 °C
(full range)
|
± 0.02 °C at 33 °C
± 0.03 °C at 350 °C |
± 0.03 °C at 50 °C
± 0.05 °C at 660 °C |
Radial uniformity
|
± 0.01 °C
(full range)
|
± 0.01 °C at 33 °C
± 0.02 °C at 350 °C |
± 0.02 °C at 50 °C
± 0.1 °C at 660 °C |
Axial uniformity
at 40 mm (1.6 in) |
± 0.05 °C
(full range)
|
± 0.04 °C at 33 °C
± 0.2 °C at 350 °C |
± 0.05 °C at 50 °C
± 0.5 °C at 660 °C |
Immersion depth
|
150 mm (5.9 in)
|
||
Heating time
(max)
|
23 min
|
5 min
|
15 min
|
Cooling time
|
15 min to –25 °C
|
14 min to 100 °C
|
25 min to 100 °C
|
Size (HWD)
|
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
|
||
Weight
|
8.2 kg (18 lbs)
|
7.3 kg (16 lbs)
|
7.7 kg (17 lbs)
|
RTD inputs
(process version) |
2-, 3-, 4-wire RTD. Resistance range of 0 to 400 ohms. Ni-120,
PT-100 (385),
PT-100 (3926), PT-100 (JIS), or Res |
||
TC inputs
(process version)
|
Type J, K, T, E, R, S, M, L, U, N, C, and mV
|
||
mA specs
(process version)
|
Range: 4-24 mA with 24-28 VDC loop power
mA Accuracy: 0.02 % of rdg + 0.002 mA |